Comparative tonometric measurements of eye bank eyes.

Abstract

A comparison was made of the accuracy of the Tono-Pen, Pneumatonometer, and Perkins hand-held tonometers by measuring the hydrostatically controlled intracular pressure from 10 to 50 mm Hg in human eye bank eyes. The open stopcock method was used in which the cannulated eye, the hydrostatic pressure controlling column, and a manometer were continuously open to each other. The Pneumatonometer gave accurate results at 50 mm Hg but overestimated the lower intraocular pressures; whereas, the Perkins gave satifactory results at 10 and 20 mm Hg but underestimated higher intraocular pressures. The Tono-Pen displayed the least deviation from the controlled pressure over the entire range studied. Another aspect of this study involved determining that a therapeutic soft contact lens did not have any noticeably adverse effect on the accuracy of the Pneumatonometer and Tono-Pen tonometers.

Authors+Show Affiliations

,

Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami School of Medicine, FL 33101.

, , , , , ,

Source

Cornea 6:3 1987 pg 219-25

MeSH

Contact Lenses, Hydrophilic
Eye Banks
Humans
Intraocular Pressure
Manometry
Ocular Physiological Phenomena
Tonometry, Ocular

Pub Type(s)

Comparative Study
Journal Article
Research Support, Non-U.S. Gov't

Language

eng

PubMed ID

3677726

Citation

Mendelsohn, A D., et al. "Comparative Tonometric Measurements of Eye Bank Eyes." Cornea, vol. 6, no. 3, 1987, pp. 219-25.
Mendelsohn AD, Forster RK, Mendelsohn SL, et al. Comparative tonometric measurements of eye bank eyes. Cornea. 1987;6(3):219-25.
Mendelsohn, A. D., Forster, R. K., Mendelsohn, S. L., Dennis, J. J., Heidemann, D. G., Levine, I. K., ... Di Maio, J. M. (1987). Comparative tonometric measurements of eye bank eyes. Cornea, 6(3), pp. 219-25.
Mendelsohn AD, et al. Comparative Tonometric Measurements of Eye Bank Eyes. Cornea. 1987;6(3):219-25. PubMed PMID: 3677726.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Comparative tonometric measurements of eye bank eyes. AU - Mendelsohn,A D, AU - Forster,R K, AU - Mendelsohn,S L, AU - Dennis,J J, AU - Heidemann,D G, AU - Levine,I K, AU - Parel,J M, AU - Di Maio,J M, PY - 1987/1/1/pubmed PY - 1987/1/1/medline PY - 1987/1/1/entrez SP - 219 EP - 25 JF - Cornea JO - Cornea VL - 6 IS - 3 N2 - A comparison was made of the accuracy of the Tono-Pen, Pneumatonometer, and Perkins hand-held tonometers by measuring the hydrostatically controlled intracular pressure from 10 to 50 mm Hg in human eye bank eyes. The open stopcock method was used in which the cannulated eye, the hydrostatic pressure controlling column, and a manometer were continuously open to each other. The Pneumatonometer gave accurate results at 50 mm Hg but overestimated the lower intraocular pressures; whereas, the Perkins gave satifactory results at 10 and 20 mm Hg but underestimated higher intraocular pressures. The Tono-Pen displayed the least deviation from the controlled pressure over the entire range studied. Another aspect of this study involved determining that a therapeutic soft contact lens did not have any noticeably adverse effect on the accuracy of the Pneumatonometer and Tono-Pen tonometers. SN - 0277-3740 UR - https://wwww.unboundmedicine.com/medline/citation/3677726/Comparative_tonometric_measurements_of_eye_bank_eyes_ DB - PRIME DP - Unbound Medicine ER -